Handheld digital manometer PG-100N

Handheld digital manometer PG-100N

 

FeaturesFeatures
  • Excellent stability

    High precision semiconductor chip combined with temperature compensation circuit guarantee excellent stability..
  • Long battery life

    Lasts more than 1000hrs with two AAA dry batteries.
  • Automatic power off function

  • Easy, one-handed operation

  • Pressure unit selectable type also available (PG-100N-****-W).

MODEL NUMBER DESIGNATION

PG-100N-102R-MX2T

Series name

Rated pressure range

101:-10~10kPa

102:-100~100kPa

103:-100~1000kPa

Pressure reference

R:Gauge(Compound)

Unit of display

Blank:kPa

M:MPa

H:mmHg 

W:Selectable✽1

Inspection report/Calibration Certificate

Blank:Not attached

X1:Inspection report

X2:nspection report and Calibration certificate

Traceability system chart

Blank:Not attached

T:Traceability system chart

 
✽1) 103R range products do not correspond to Unit mmHg of Pressure unit selectable W type. 
(Refer to the Pressure unit selectable W type below.)

Pressure unit selectable W type

Pressure rangeUnit selectable W type
kPaMpammHgin.HgmmAqpsibar
101R
102R
103R

※ " ー" is not supported as a display unit.

LIST OF MODEL NUMBERS

Pressure reference Gauge(Compound)
Rated pressure range (kPa)-10~10-100~100-100~1000
kPa Product only PG-100N-101R PG-100N-102R PG-100N-103R
Inspection report PG-100N-101R-X1 PG-100N-102R-X1 PG-100N-103R-X1
Inspection report and Traceability system chart PG-100N-101R-X1T PG-100N-102R-X1T PG-100N-103R-X1T
Inspection report and Calibration certificate PG-100N-101R-X2 PG-100N-102R-X2 PG-100N-103R-X2
Traceability system chart PG-100N-101R-T PG-100N-102R-T PG-100N-103R-T
Inspection report,Calibration certificate and Traceability system chart PG-100N-101R-X2T PG-100N-102R-X2T PG-100N-103R-X2T
MPa Product only ――――  ――――   PG-100N-103R-M
Inspection report PG-100N-103R-M-X1
Inspection report and Traceability system chart PG-100N-103R-M-X1T
Inspection report and Calibration certificate PG-100N-103R-M-X2
Traceability system chart PG-100N-103R-M-T
Inspection report,Calibration certificate and Traceability system chart PG-100N-103R-M-X2T
mmHg Product only PG-100N-101R-H PG-100N-102R-H ――――  
Inspection report PG-100N-101R-H-X1 PG-100N-102R-H-X1
Inspection report and Traceability system chart PG-100N-101R-H-X1T PG-100N-102R-H-X1T
Inspection report and Calibration certificate PG-100N-101R-H-X2 PG-100N-102R-H-X2
Traceability system chart PG-100N-101R-H-T PG-100N-102R-H-T
Inspection report,Calibration certificate and Traceability system chart PG-100N-101R-H-X2T PG-100N-102R-H-X2T
Selectable Product only  PG-100N-101R-W PG-100N-102R-W PG-100N-103R-W
Inspection report PG-100N-101R-W-X1 PG-100N-102R-W-X1 PG-100N-103R-W-X1
Inspection report and Traceability system chart PG-100N-101R-W-X1T PG-100N-102R-W-X1T PG-100N-103R-W-X1T
Inspection report and Calibration certificate PG-100N-101R-W-X2 PG-100N-102R-W-X2 PG-100N-103R-W-X2
Traceability system chart PG-100N-101R-W-T PG-100N-102R-W-T PG-100N-103R-W-T
Inspection report,Calibration certificate and Traceability system chart PG-100N-101R-W-X2T PG-100N-102R-W-X2T PG-100N-103R-W-X2T

 

Standard specifications[kPa version]

◆Unless otherwise specified, the specs are defined at an ambient temperature of 25 ± 5 °C.
ItemPG-100N

General
specifications

Pressure reference Gauge
Model number 101R 102R 103R
Rated pressure range −10 ~ 10kPa −100 ~ 100kPa −100 ~ 1000kPa
Maximum pressure 20kPa 200kPa 1500kPa
Break-down pressure 50kPa 500kPa 2000kPa
Pressure medium Non-corrosive gases
Operating temp. range 0 ~ 50°C
Compensated temp. range 0 ~ 50°C
Operating humidity 15 ~ 85%RH(No condensation)
Storage temp. −20 ~ 70°C(Atmospheric pressure, humidity 65 %RH maximum)
Pressure port M 5 female
Net weight Approx. 100g
Power AAA dry battery DC 1.5 V × 2 Battery life 1000 h minimum

Display

No. of digits 3 1/2
Display element LCD
Rated display range −10.00 ~ 10.00kPa −100.0 ~ 100.0kPa −100 ~ 1000 kPa/−. 100 ~ 1.000(MPa version)
Display cycle Approx. 4 times/s
Integrated accuracy ± 0.5 %F.S. ± 2 digits (at 25 °C ± 5 °C)
± 1.5 %F.S. ± 2 digits
(at 25 °C ± 25 °C)
± 1.0 %F.S. ± 2 digits
(at 25 °C ± 25 °C)
Resolution 1 digit

Analog
output

Power supply
(External supply)
5 ± 0.5 V DC 5 mA maximum
Output voltage 0 ~ ± 1.000 V ※Proportionate to display value
Output accuracy/Resolution ± 10 mV / 1mV
Output resistance Approx. 10 k Ω
Response Approx. 10 ms
 Additional function Auto power off, Display hold, Battery replacement display, (LOBAT), Zero point adjustment
Accessories 2 fittings (half union for 6mm O.D. tube, barb fitting for 4mm I.D. tube), 2 AAA batteries, strap
Option
(Sold seperately)
PG-100N soft carrying case

Standard specifications[Non SI unit version]

◆Unless otherwise specified, the specs are defined at an ambient temperature of 25±5 °C.

 ItemPG-100N

General
specifications

Pressure reference Gauge
Model number
(mmHg display : compound)
101R-H 102R-H
Rated pressure range
kPa(mmHg)
− 10 ~ 10(− 75.0 ~ 75.0) − 100 ~ 100(− 750 ~ 750)
Maximum pressure
kPa(mmHg)
20kPa (150) 200(1500)
Break-down pressure
kPa(mmHg)
50kPa (375) 500(3750)
Pressure medium Non-corrosive gases
Operating temp. range 0 ~ 50°C
Compensated temp. range 0 ~ 50°C
Operating humidity 15 ~ 85%RH(No condensation)
Storage temp. −20 ~ 70°C(Atmospheric pressure, humidity 65 %RH maximum)
Pressure port M 5 female
Net weight Approx. 100g
Power AAA dry battery DC 1.5 V × 2 Battery life 1000 h minimum
 

Display

No. of digits 3 1/2
Display element LCD
Rated display range
(mmHg display)
− 75.0 ~ 75.0 − 750 ~ 750
Display cycle Approx. 4 times/s
Integrated accuracy ± 0.5 %F.S. ± 2 digits (at 25 °C ± 5 °C)
± 1.5 %F.S. ± 2 digits
(at 25 °C ± 25 °C)
± 1.0 %F.S. ± 2 digits
(at 25 °C ± 25 °C)
Resolution 1 digit

Analog
output

 
Power supply
(External supply)
5 ± 0.5 V DC 5 mA maximum
Output voltage 0 ~ ± 1.000 V ※Proportionate to display value
Output accurac/Pesolution ± 10 mV / 1mV
Output resistance Approx. 10 kΩ
Response Approx. 10 ms
   Additional function Auto power off , Display hold,
Battery replacement display(LOBAT),
Zero point adjustment
Accessories 2 fittings (half union for 6mm O.D. tube, barb fitting for 4mm I.D. tube),
2 AAA batteries, strap
Option
(Sold seperately)
PG-100N soft carrying case

ENVIRONMENTAL CHARACTERISTICS

Test itemTest conditions(At 25 ± 5 °C)Permissible change
Vibration Amplitude 1.5 mm P-P, 10 ~ 55 Hz,
3 directions for 2 hours each (packed)
Indicating accuracy: ± 1 %F.S. maximum


Shock 98.1 m/s2 , 3 directions for 3 times each (packed)
Moisture resistance 40 °C, 90 ~ 95 %RH, 240 hrs
Pressure cycling 0 ~ Rated pressure, 106 cycles

OUTLINE DIMENSIONS

Unless otherwise specified, tolerance : ± 0.5 (Unit: mm)

Troubleshooting

Symptom/ProblemPossible causeAction/Troubleshooting
Fluctuation in display value. Fluctuation in display value. Fluctuation in measured pressure or leakage.
  • Check for air leaks from the attachments or pipe joints.
  • If no problems are found on the use environment, please contact our sales team.
Zero point deviation Zero point deviation Residual pressure in the pipework.
  • When measuring guage pressure, please make sure that pipework has open access to atmospheric pressure
  • Please set the zero point when the pipe has open access to the atmophere.
  • Should you be measuring absolute pressure then please contact our sales team.
  • For other display deviations, recalibration is required, please contact our sales team.。
No display when power is on. No display when power is on. Reversed polarity of dry batteries.
  • Check the polarity of the dry batteries.
  • If the polarity is correct then replace with new batteries
  • Commercially available AAA batteries are used in the PG-100B.
  • If the system is not restored after replacing the batteries, please contact our sales team.
Display shows "LOBAT". Display shows "LOBAT". Low battery.
  • Replace the battery when "LOBAT" is shown on the display.
  • Please replace the two AAA dry batteries and dispose the used batteries in accordance with the instructions of the local government.
  • Make a zero-adjustment after replacing the batteries.
Flashing display. Flashing display. A pressure exceeding 130% of the rating is applied.
  • Check the source pressure and use within the rated pressure range.
●The above contents and descriptions are subject to change without notice. 