Liquid leak sensor WL10
- ●To make a purchase, contact Distributors, or use e-commerce below.
SALES CHANNEL ;Check our Distributors.- SALES CHANNELClick here to check our Distributors.
Liquid leak sensor WL10
- Operable in various area thanks to unique optical technology
- Can detect leakage without a douser and additional amplifying circuit
- PFA casing available for detecting leakage of corrosive liquid
- Compact and Slim size ( φ 25 mm, t = 10 mm)
- Lead-free Soldering,UL standard,CE Marking
General specifications
- Unless otherwise specified, the specs are defined at an ambient temperature of 25±5 °C and excitation voltage of 24 VDC.
Operating temp. range | -10~60°C |
---|---|
Operating humidity | 35 ~ 85%RH(No condensation) |
Storage temp. | -20~70°C(Atmospheric pressure, humidity 65 %RH maximum) |
Pressure medium | Water (PP) / Corrosive fluids compatible with PFA |
Material of housing | PP/PFA |
Net weight |
Approx. 30g(WL10- * P-2) |
Supply voltage | 12~24±10 %VDC |
Consumption current | 10mA max. (Except open collector output) |
Output
- Unless otherwise specified, the specs are defined at an ambient temperature of 25±5 °C and excitation voltage of 24 VDC.
Switch output interface | Switching capacity | Residual voltage | Operation grade |
---|---|---|---|
NPN | Voltage endurance : 30 VDC Sink current : 50 mA maximum (Over current protection) |
1.0 V maximum | Normal condition : Switch output is ON / Green LED is ON Leakage detection : Switch output is OFF / Red LED is ON |
PNP | Voltage endurance : 24 VDC Sourced current : 50 mA maximum (Over current protection) |
2.0 V maximum | Normal condition : Switch output is ON / Green LED is ON Leakage detection : Switch output is OFF / Red LED is ON |
ENVIRONMENTAL CHARACTERISTICS
Test item | Test conditions | Specification |
---|---|---|
Vibration | 10~500 Hz, 1.5 mm maximum,98.1 m/s2 , 3 directions for 2 hours each |
Meets standard specifications |
Shock | 490 m/s2 , 3 directions for 3 times each | |
Protection grade | IP67 |
MODEL NUMBER DESIGNATION
WL10- | N | P- | 2 |
---|---|---|---|
Series name |
Switch output interface N:NPN open collector P:PNP open collector |
Material of housing P:PP(Polypropylene) F:PFA(Perfluoroalkoxy alkane) |
Cable length 2:2 m 4:4 m |
LIST OF MODEL NUMBERS
Model number | Switch output interface | Material of housing | Cable length | CAD |
---|---|---|---|---|
WL10-NP-2 | NPN | PP | 2 m | |
WL10-NP-4 | NPN | PP | 4 m | |
WL10-NF-2 | NPN | PFA | 2 m | |
WL10-NF-4 | NPN | PFA | 4 m | |
WL10-PP-2 | PNP | PP | 2 m | |
WL10-PP-4 | PNP | PP | 4 m | |
WL10-PF-2 | PNP | PFA | 2 m | |
WL10-PF-4 | PNP | PFA | 4 m |
INTERNAL ELECTRICAL SCHEMATICS
OUTLINE DIMENSION
Unless otherwise specified, tolerance: ± 0.5 (Unit: mm)
Documents
- ●The above contents and descriptions are subject to change without notice.