Handheld digital manometer PG-100B

●To buy this product, please go to the following Online Web stores.
To the Handheld digital manometer “PG-100B series” page on the DigiKey online shop
To the Handheld digital manometer “PG-100B series” page on the Mouser online shop
To the Handheld digital manometer “PG-100B series” page on the RS components online shop

Handheld digital manometer PG-100B

 

FeaturesFeatures
  • New lineup of PG-100 series for measuring gas and liquid pressure

  • Capable of measuring Absolute pressure (PG-100B-102A)

  • Excellent stability
    High precision semiconductor chip combined with temperature compensation circuit guarantee excellent stability.

  • Long battery life 

    Lasts more than 1000hrs with two AAA dry batteries.
  • Automatic power off function

  • Easy, one-handed operation

  • Easy-to-read large LCD display

  • Pressure unit selectable type also available (PG-100B-****-W)

Standard specifications[kPa version]

  • Unless otherwise specified, the specs are defined at an ambient temperature of 25±5 °C.
General specificationsPressure reference Gauge Absolute
Model number 102R 103R 102A
Rated pressure range −100~100kPa −100~1000kPa 0~100 kPa abs
Maximum pressure 200kPa 2000kPa 200 kPa abs
Break-down pressure 300kPa 3000kPa 300 kPa abs
Pressure medium Gases/liquids compatible with SUS316L
Operating temp. range 0 ~ 50°C
Compensated temp. range 0 ~ 50°C
Operating humidity 15 ~ 85%RH(No condensation)
Storage temp. −20 ~ 70°C(Atmospheric pressure, humidity 65 %RH maximum)
Pressure port MM 5 female
Net weight Approx. 100g
Power AAA dry battery DC 1.5 V × 2 Battery life 1000 h minimum
DisplayNo. of digits 3 1/2
Display element LCD
Rated display range -100.0 ~ 100.0 kPa −100 ~ 1000 kPa
−. 100 ~ 1.000(MPa version)
00.0 ~ 100.0 kPa
Display cycle Approx. 4 times/s
Integrated accuracy ± 0.5 %F.S. ± 2 digits
(at 25 °C ± 5 °C)
± 1.0 %F.S. ± 2 digits
(at 25 °C ± 25 °C)
± 2.0 %F.S. ± 2 digits
(at 25 °C ± 25 °C)
Resolution 1 digit
Analog outputPower supply
(External supply)
5 ± 0.5 V DC 5 mA maximum
Output voltage 0 ~ ± 1.000 V ※ Proportionate to display value
Output accuracy/Resolution ± 10 mV / 1mV
Output resistance Approx. 10 kΩ
Response Approx. 10 ms
   Additional function Auto power off, Display hold,
Battery replacement display(LOBAT),
Zero point adjustment
Accessories Fitting (one touch for 6mm O.D. tube),Insert ring,
2 AAA batteries, strap, protective cover
Option
(Sold seperately)
PG-100B soft carrying case

Standard specifications[Non SI unit version]

※Unless otherwise specified, the specs are defined at an ambient temperature of 25±5 °C.

General
specifications

Pressure reference Gauge
Model number
(mmHg display:compound)
102R-H
Rated pressure range -100~100 (-750 ~ 750) kPa(mmHg)
Maximum pressure 200(1500) kPa(mmHg)
Break-down pressure 300 (2250) kpa(mmHg)
Operating temp. range 0~50 °C
Compensated temp. range 0~50 °C
Operating humidity 35~85% RH(No condensation)
Storage temp. -20~70°C
(Atmospheric pressure, humidity 65 %RH maximum)
Pressure medium Gases/liquids compatible with SUS316L
Pressure port M 5 female
Net weight Approx. 100g
Power AAA dry battery DC 1.5 V ×
2 Battery life 1000 h minimum

Display

No. of digits 3 1/2
Display element LCD
Rated display range
(mmHg display)
− 750 ~ 750
Display cycle Approx. 4 times/s
Resolution ± 0.5 %F.S. ± 2 digits
(at 25 °C ± 5 °C)
± 1.0 %F.S. ± 2 digits
(at 25 °C ± 25 °C)
Resolution 1 digit

Analog output

Power supply
(External supply)
5 ± 0.5 V DC 5 mA maximum
Output voltage 0 ~ ± 1.000 V  ※Proportionate to display value
Output accuracy/Pesolution ± 10 mV / 1 mV
Output resistance Approx. 10 kΩ
Response Approx. 10 ms
 Additional function Auto power off , Display hold,
Battery replacement display(LOBAT), Zero point adjustment
Accessories Fitting (one touch for 6mm O.D. tube),Insert ring,
2 AAA batteries, strap, protective cover
Option
(Sold seperately)
PG-100B soft carrying case

ENVIRONMENTAL CHARACTERISTICS

Test itemTest conditions(At 25 ± 5 °C)Permissible change
Vibration Amplitude 1.5 mm P-P, 10 ~ 55 Hz,
3 directions for 2 hours each (packed)
Indicating accuracy: ± 1 %F.S. maximum

Shock 98.1 m/s2 , 3 directions for 3 times each (packed)
Moisture resistance 40 °C, 90 ~ 95 %RH, 240 hrs
Pressure cycling 0 ~ Rated pressure, 106cycles

MODEL NUMBER DESIGNATION

PG-100B-102RMX2T

Series name

Rated pressure range/Pressure reference 

102R:-100~100 kPa/

           Gauge(Compound)

103R:-100~1000 kPa/

           Gauge(Compound)

102A:0~100 kPa abs/

           Absolute

Unit of display

Blank:kPa

M:MPa

H:mmHg 

W:Selectable✽1

Inspection report/Calibration Certificate

Blank:Not attached

X1:Inspection report

X2:Inspection report and

      Calibration certificate

Traceability system chart

Blank:Not attached

 T:Traceability system chart

 
✽1) 103R range products do not correspond to Unit mmHg of Pressure unit selectable W type. 
(Refer to the Pressure unit selectable W type below.)

Pressure unit selectable W type

Pressure rangeUnit selectable W type
kPaMpammHgin.HgmmAqpsibar
101R
102R
103R

※ " ー" is not supported as a display unit.

LIST OF MODEL NUMBERS

Pressure reference Gauge(Compound)Absolute
Rated pressure range (kPa)-100~100-100~10000 ~ 100(abs)
kPa Product only PG-100B-102R PG-100B-103R PG-100B-102A
Inspection report PG-100B-102R-X1 PG-100B-103R-X1 PG-100B-102A-X1
Inspection report and Traceability system chart PG-100B-102R-X1T PG-100B-103R-X1T PG-100B-102A-X1T
Inspection report and Calibration certificate PG-100B-102R-X2 PG-100B-103R-X2 PG-100B-102A-X2
Traceability system chart PG-100B-102R-T PG-100B-103R-T PG-100B-102A-T
Inspection report,Calibration certificate and Traceability system chart PG-100B-102R-X2T PG-100B-103R-X2T PG-100B-102A-X2T
MPa Product only  ――――――― PG-100B-103R-M ―――――――
Inspection report PG-100B-103R-M-X1
Inspection report and Traceability system chart PG-100B-103R-M-X1T
Inspection report and Calibration certificate PG-100B-103R-M-X2
Traceability system chart PG-100B-103R-M-T
Inspection report,Calibration certificate and Traceability system chart PG-100B-103R-M-X2T
mmHg Product only PG-100B-102R-H ――――――― ―――――――
Inspection report PG-100B-102R-H-X1
Inspection report and Traceability system chart PG-100B-102R-H-X1T
Inspection report and Calibration certificate PG-100B-102R-H-X2
Traceability system chart PG-100B-102R-H-T
Inspection report,Calibration certificate and Traceability system chart PG-100B-102R-H-X2T
Selectable Product only PG-100B-102R-W PG-100B-103R-W PG-100B-102A-W
Inspection report PG-100B-102R-W-X1 PG-100B-103R-W-X1 PG-100B-102A-W-X1
Inspection report and Traceability system chart PG-100B-102R-W-X1T PG-100B-103R-W-X1T PG-100B-102A-W-X1T
Inspection report and Calibration certificate PG-100B-102R-W-X2 PG-100B-103R-W-X2 PG-100B-102A-W-X2
Traceability system chart PG-100B-102R-W-T PG-100B-103R-W-T PG-100B-102A-W-T
Inspection report,Calibration certificate and Traceability system chart PG-100B-102R-W-X2T PG-100B-103R-W-X2T PG-100B-102A-W-X2T

OUTLINE DIMENSIONS

Unless otherwise specified, tolerance : ± 0.5 (Unit: mm)

Troubleshooting

Symptom/ProblemPossible causeAction/Troubleshooting
Fluctuation in display value. Fluctuation in display value. Fluctuation in measured pressure or leakage.
  • Check for air leaks from the attachments or pipe joints.
  • If no problems are found on the use environment, please contact our sales team.
Zero point deviation Zero point deviation Residual pressure in the pipework.
  • When measuring guage pressure, please make sure that pipework has open access to atmospheric pressure
  • Please set the zero point when the pipe has open access to the atmophere.
  • Should you be measuring absolute pressure then please contact our sales team.
  • For other display deviations, recalibration is required, please contact our sales team.。
No display when power is on. No display when power is on. Reversed polarity of dry batteries.
  • Check the polarity of the dry batteries.
  • If the polarity is correct then replace with new batteries
  • Commercially available AAA batteries are used in the PG-100B.
  • If the system is not restored after replacing the batteries, please contact our sales team.
Display shows "LOBAT". Display shows "LOBAT". Low battery.
  • Replace the battery when "LOBAT" is shown on the display.
  • Please replace the two AAA dry batteries and dispose the used batteries in accordance with the instructions of the local government.
  • Make a zero-adjustment after replacing the batteries.
Flashing display. Flashing display. A pressure exceeding 130% of the rating is applied.
  • Check the source pressure and use within the rated pressure range.
How to mount the attachment How to mount the attachment  
  • Please refer to "PG-100B Attachment"below.

 

How to mount the attachment

1. One Touch Fitting
 One Touch Fitting(※With washer)
    • [Manufacturer / Model]
      NIHON PISCO CO., LTD / SSC6-M5
    • [Material]
      Metal parts: SUS316, Seal Rubber: FKM
  • [Specification]
    Fluid medium: Air, Water (Conditional*) Others such as chemicals (Conditional*)
    Max. operating pressure: 1.0 MPa
    Max. vacuum: -100 kPa
    Operating temp. range: -15~120 °C (No freezing)

 [Dimensions]

(Unit: mm)  
Tube φD 6 φP 11.8
L 22.6 R M5×0.8
A 3 H (Opposite side) 12
B 25.6 Effective cross-sectional area 3 mm2
C 19.6 Weight 8.4 g

 

⚠️ Attention⚠️ Attention
Make sure to follow the instructions below when the fluid medium is water or other chemicals.
1.Surge pressure must be controlled lower than max. operating pressure when using water or liquid as a fluid medium.
2.Be sure to place Insert Ring into the tube edge when the tube is inserted into Push-In fittings with water liquid as a fluid medium. (please see below)
3.The specification above may not be applied, depending on the kind of chemicals or mixed gases used as fluid medium.
 
(1) How to insert and to detach a tube

❶ How to insert a tube

  • Simpry insert a tube to push-in tube fitting and it is completed.
    Make sure to insert a tube up to tube-end.
    Lock-claws bite a tube and elastic sleeve seals around the tube.
  • When installing, please refer to the C dimension in the upper dimension table.

❷ How to detach a tube

  • Pull out a tube with pushing release-ring, and the tube gets detached.
  • Make sure to stop air supply before you detach a tube.
 
(2) How to tighten the screws

❶ Tighten the screws

  • There are two types of tightening screws: one is the method of tightening the outer diameter hexagon with a wrench, and the other is the method of tightening the hexagon inner diameter with a hexagonal wrench (you can leave a small space between the joints).
  • In tightening the screw, refer to the following Note 2.
Note 1. Make sure that the outside diameter tolerance of the tube meets the following specifications.
MM size (mm)Nylon tubeUrethane tube
 φ6mm  ±0.1mm  ±0.15mm
Note 2. When tightening screws, please refer to the following recommended tightening torque.
Screw typeScrew sizeTightening torque
Metric thread  M5×0.8  1.0~1.5N・m
 
2. Insert Ring
 Insert Ring
 
  [Manufacturer / Model]
NIHON PISCO CO., LTD / SSC6-M5

  [Dimensions]

(Unit: mm) 
Applicable tubeD1D2BThicknessWeight (g)
O.D.φ6/I.D.φ4 φ4 φ4.5 18.5 0.2  0.35
 
External Installation

 

⚠️ Attention⚠️ AttentionWhen using liquid as a medium or connecting the extremely soft tube to the one-touch fitting, be sure to use the included insert ring. If not used, the tube may fall off.
●The above contents and descriptions are subject to change without notice. 

Products

×